HomeProductsIsolated energy distributionMedical IT Isolated Power SystemAIM-M200 Medical Intelligent Isolation Monitor

AIM-M200 Medical Intelligent Isolation Monitor

The AIM-M200 intelligent medical insulation monitoring device utilizes advanced microcontroller technology, featuring high integration, compact size, and easy installation, and combines intelligence, digitalization, and networking. It is ideal for insulation monitoring of isolated power systems in Class 2 medical environments such as operating rooms and intensive care units.

  • Isolation monitoring for medical IT systems
  • Load and temperature monitoring of the isolation transformer
  • Device disconnection monitoring
  • LED indicators: On, Isolation, Overload, Overtemperature
  • Alarm relay output
  • Test button
  • Insulation fault location
  • RS485 (MODBUS RTU) + CAN

Product info

function
1) Real-time monitoring and fault alarm of earth insulation resistance,
2) Display of the transformer load current and the transformer winding temperature of the monitored IT system,
3) Use with an insulation fault locator that starts fault location remotely and displays the location results for insulation faults,
4) Real-time monitoring of line open faults, temperature sensor open faults and functional earth line open faults of the monitored system with alarm notification within 2 seconds after the fault occurs,
5) Relay alarm output, LED alarm indicator and other fault display functions,
6) Use as a central alarm and display device, test signal generator, insulation fault locator and communication with higher-level computer management software for real-time monitoring of the operating status of the IT system.
7) Event logging: Records the alarm time and error type. This allows operators to analyze system operation and promptly resolve errors.

Technical data

AUX power supply

voltage

DC 18-36V

Temperature monitoring

Thermal resistance

2 Pt100

Power consumption

≤3VA

measuring range

-50~+200℃

Insulation monitoring

Resistance measuring range

10-999kΩ

Alarm value range

0~+200℃

Response value

50-999kΩ

Alarm output

Output mode

2 relays

Relative uncertainty

±10%

Contact rating

AC 250V/3A

DC 30V/3A

response time

≤3s

Environment

operating temperatur

-10~+55℃

Permissible system leakage capacitance Ce

≤5uF

Transport temperature

-25~+70℃

Measuring voltage Um

≤12V

Storage temperature

-25~+70℃

Measuring current Im

≤50uA

Relative humidity

5%-95%, No condensation

Impedance Zi

≥200kΩ

Height

≤2500m

Internal DC resistance Ri

≥240kΩ

IP grade

IP30

Permissible external DC voltage Ufg

≤DC280V

Rated impulse voltage / pollution degree

4KV/III

Load current monitoring

Measured value

2.1-50A

EMC

IEC 61326-2-4

Alarm value

5-50A

communication

CAN

Measurement accuracy

±5%

RS485(Modbus RTU)