AIM-M100 Medical Isolation Monitor
The Acrel AIM-M100 medical insulation resistance tester measures the insulation resistance of medical IT systems. It also simultaneously monitors the load current and temperature of the IT system transformer. It can be used in combination with ASG100 and AIL100 series insulation fault locators and the corresponding measuring current transformers.
Product info

Technical data
AUX power supply |
voltage |
DC 18-36V |
Temperature monitoring |
Thermal resistance |
2 Pt100 |
Power consumption |
≤3VA |
measuring range |
-50~+200℃ |
||
Insulation monitoring |
Resistance measuring range |
10-999kΩ |
Alarm value range |
0~+200℃ |
|
Response value |
50-999kΩ |
Alarm output |
Output mode |
2 relays |
|
Relative uncertainty |
±10% |
Contact rating |
AC 250V/3A DC 30V/3A |
||
response time |
≤3s |
Environment |
operating temperatur |
-10~+55℃ |
|
Permissible system leakage capacitance Ce |
≤5uF |
Transport temperature |
-25~+70℃ |
||
Measuring voltage Um |
≤12V |
Storage temperature |
-25~+70℃ |
||
Measuring current Im |
≤50uA |
Relative humidity |
5%-95%, No condensation |
||
Impedance Zi |
≥200kΩ |
Height |
≤2500m |
||
Internal DC resistance Ri |
≥240kΩ |
IP grade |
IP30 |
||
Permissible external DC voltage Ufg |
≤DC280V |
Rated impulse voltage / pollution degree |
4KV/III |
||
Load current monitoring |
Measured value |
2.1-50A |
EMC |
IEC 61326-2-4 |
|
Alarm value |
5-50A |
communication |
CAN |
||
Measurement accuracy |
±5% |
RS485(Modbus RTU) |